Calibration method of conformal X-ray transmission imaging system based on small ball phantom

Yang LI, Shaohua ZHI, Xuanqin MOU

Research output: Book Chapters | Papers in Conference ProceedingsConference paper (refereed)Researchpeer-review

Abstract

Compared to conventional X-ray transmission imaging techniques, imaging systems based on field-emissive cathode-based conformal transmission imaging schemes have smaller volumes. This imaging system combines multiple point sources to complete very low dose imaging of the region of interest using a small cone beam, and finally obtain high-quality images by image restoration methods. However, more precise system geometry is required by the system architecture of the new imaging scheme and the special image-forming method. If the geometric parameters that are inconsistent with the actual use are used in image restoration, it will cause the geometric distortion in the final restored image. In this work, a proposed geometric calibration applied to the new imaging scheme. First, we use a partial point source on the periphery of the array ray source to project the side ball phantom of the imaged object. Then, a cost function that associates the geometrical parameters with the degree to which the back-projections of the ball phantom in projections from different point sources is constructed. Finally, the particle swarm optimization (PSO) is used for minimizing the cost function, and the resultant value of the cost function is the actual geometric parameter. With this calibration method, transmission imaging and geometric calibration will be done simultaneously. Besides, this calibration method does not depend on a complex calibration phantom and is simple and effective. In addition, the PSO can accelerate the implementation of the algorithm. The effectiveness and accuracy of this calibration method are verified by two sets of simulation experiments.

Original languageEnglish
Title of host publicationMedical Imaging 2020: Physics of Medical Imaging
EditorsGuang-Hong CHEN, Hilde BOSMANS
PublisherSPIE
ISBN (Electronic)9781510633919
DOIs
Publication statusPublished - 2020
Externally publishedYes
EventMedical Imaging 2020: Physics of Medical Imaging - Houston, United States
Duration: 16 Feb 202019 Feb 2020

Publication series

NameProgress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume11312
ISSN (Print)1605-7422

Conference

ConferenceMedical Imaging 2020: Physics of Medical Imaging
Country/TerritoryUnited States
CityHouston
Period16/02/2019/02/20

Bibliographical note

Publisher Copyright:
© 2020 SPIE.

Keywords

  • Conformal Imaging
  • Geometric Calibration
  • Particle Swarm Optimization (PSO)
  • Transmission Imaging

Fingerprint

Dive into the research topics of 'Calibration method of conformal X-ray transmission imaging system based on small ball phantom'. Together they form a unique fingerprint.

Cite this